Tag | (0028,0300) |
---|---|
Type | Optional (3) |
Keyword | QualityControlImage |
Value Multiplicity | 1 |
Value Representation | Code String (CS) |
Example Values |
|
Indicates whether or not this image is a quality control or phantom image.
Enumerated Values:
If this Attribute is absent, then the image may or may not be a quality control or phantom image. The phantom device in the image can be described using the Device Module. See Section C.7.6.12
Table C.7-18 describes the Attributes of devices or calibration objects (e.g., catheters, markers, baskets) that are associated with a Study and/or image.
Table C.7-18. Device Module Attributes
Attribute Name |
Tag |
Type |
Attribute Description |
---|---|---|---|
Device Sequence |
(0050,0010) |
1 |
A Sequence of Items describing devices used that may be visible in the image. One or more Items shall be included in this Sequence. |
>Manufacturer |
(0008,0070) |
3 |
Manufacturer of the device |
>Manufacturer's Model Name |
(0008,1090) |
3 |
Manufacturer's model name of the device |
>Device Serial Number |
(0018,1000) |
3 |
Manufacturer's serial number of the device |
>Device ID |
(0018,1003) |
3 |
User-supplied identifier for the device |
>Device Length |
(0050,0014) |
3 |
Length in mm of device. See Section C.7.6.12.1.1. |
>Device Diameter |
(0050,0016) |
3 |
Unit diameter of device. See Section C.7.6.12.1.1. |
>Device Diameter Units |
(0050,0017) |
2C |
Required if Device Diameter (0050,0016) is present. Defined Terms:
|
>Device Volume |
(0050,0018) |
3 |
Volume of device in ml. See Section C.7.6.12.1.1. |
>Inter-Marker Distance |
(0050,0019) |
3 |
Distance in mm between markers on calibrated device. See Section C.7.6.12.1.1. |
>Device Description |
(0050,0020) |
3 |
Further description in free form text describing the device. |
Depending on the type of device specified by Code Value (0008,0100) in an Item of Device Sequence (0050,0010), various device size Attributes (e.g., Device Length (0050,0014), Device Diameter (0050,0016), Device Volume (0050,0018), Inter Marker Distance (0050,0019)) may be required to fully characterize the device.
Table C.7-18 describes the Attributes of devices or calibration objects (e.g., catheters, markers, baskets) that are associated with a Study and/or image.
Table C.7-18. Device Module Attributes
Attribute Name |
Tag |
Type |
Attribute Description |
---|---|---|---|
Device Sequence |
(0050,0010) |
1 |
A Sequence of Items describing devices used that may be visible in the image. One or more Items shall be included in this Sequence. |
>Manufacturer |
(0008,0070) |
3 |
Manufacturer of the device |
>Manufacturer's Model Name |
(0008,1090) |
3 |
Manufacturer's model name of the device |
>Device Serial Number |
(0018,1000) |
3 |
Manufacturer's serial number of the device |
>Device ID |
(0018,1003) |
3 |
User-supplied identifier for the device |
>Device Length |
(0050,0014) |
3 |
Length in mm of device. See Section C.7.6.12.1.1. |
>Device Diameter |
(0050,0016) |
3 |
Unit diameter of device. See Section C.7.6.12.1.1. |
>Device Diameter Units |
(0050,0017) |
2C |
Required if Device Diameter (0050,0016) is present. Defined Terms:
|
>Device Volume |
(0050,0018) |
3 |
Volume of device in ml. See Section C.7.6.12.1.1. |
>Inter-Marker Distance |
(0050,0019) |
3 |
Distance in mm between markers on calibrated device. See Section C.7.6.12.1.1. |
>Device Description |
(0050,0020) |
3 |
Further description in free form text describing the device. |
Depending on the type of device specified by Code Value (0008,0100) in an Item of Device Sequence (0050,0010), various device size Attributes (e.g., Device Length (0050,0014), Device Diameter (0050,0016), Device Volume (0050,0018), Inter Marker Distance (0050,0019)) may be required to fully characterize the device.