Tag | (0018,9379) |
---|---|
Type | Required (1) |
Keyword | MultienergyCTPathSequence |
Value Multiplicity | 1 |
Value Representation | Sequence (SQ) |
Describes the paths corresponding to each energy level of a Multi-energy acquisition. Each path consists of a source (see Section C.8.2.1.1.1) and a detector (see Section C.8.2.2.1.1).
Two or more Items shall be included in this Sequence.
See also Section C.8.2.2.1 and Section C.8.2.2.2.
For CT Images, Image Type (0008,0008) is specified to be Type 1.
Defined Terms for Value 3:
a CT Cross-sectional Image
a CT Localizer Image
Cross-sectional images include transverse, coronal, sagittal and oblique images.
Image Type (0008,0008) Value 4 shall be present if Multi-energy CT Acquisition (0018,9361) has a value of YES.
Defined Terms for Value 4 for Multi-energy CT Images:
a Virtual Monoenergetic Image. Each real-world value mapped pixel represents CT Hounsfield units and is analogous to a CT image created by a monoenergetic (of a specific keV value) X-Ray beam.
a Material-Specific Image. Each real-world value mapped pixel value represents a property of a material such as attenuation, concentration or density.
An image with the attenuation contribution of one or more materials removed. For pixels that did not contain any of the removed material(s), the pixel values are unchanged.
a Material-Fractional Image. Each real-world value mapped pixel represents the fraction of a voxel occupied by a material.
an Effective Atomic Number Image. Each real-world value mapped pixel represents Effective Atomic Number of the materials in the voxel.
an Electron Density Image. Each real-world value mapped pixel represents the number of electrons per unit volume or the electron density relative to water.
a Material-Modified Image. CT Image where real-world value mapped pixels have been modified to highlight a certain target material (either by partially suppressing the background or by enhancing the target material), or to partially suppress the target material.
a Value-Based Image. CT Image where real-world value mapped pixels represent a certain value for a specified material
Multi-energy CT images are not necessarily DERIVED and may be ORIGINAL\PRIMARY.
When an image is created by a generic transformation an implementation-specific Value 4 may be provided.
Each item in the Multi-energy CT X-Ray Source Sequence (0018,9365) might describe either a constant source, a constant source with energy selective filter or one output corresponding to a specific energy of a KV switching source. The Attributes will refer to a source, meaning one item in the Multi-energy CT X-Ray Source Sequence (0018,9365). The Attributes use the phrase "physical X-Ray Source" when it is necessary to refer to the actual device (tube) rather than the current item.
Table C.8.2.2-2 specifies the Attributes of the Multi-energy CT X-Ray Source Macro, which describe the CT X-Ray Source(s).
Table C.8.2.2-2. Multi-energy CT X-Ray Source Macro Attributes
Attribute Name |
Tag |
Type |
Attribute Description |
---|---|---|---|
Multi-energy CT X-Ray Source Sequence |
(0018,9365) |
1 |
X-Ray Source (see Section C.8.2.2.1.1) information. One or more Items shall be included in this Sequence. |
>X-Ray Source Index |
(0018,9366) |
1 |
Identification number of this item in the Multi-energy CT X-Ray Source Sequence. The number shall be 1 for the first Item and increase by 1 for each subsequent Item. |
>X-Ray Source ID |
(0018,9367) |
1 |
Identifier of the physical X-Ray source. This might be the serial number. The X-Ray Source ID (0018,9367) will have the same value for different values of X-Ray Source Index (0018,9366) if a single source generates different nominal energies. |
>Multi-energy Source Technique |
(0018,9368) |
1 |
Technique used to acquire Multi-energy data. Defined Terms:
|
>Source Start DateTime |
(0018,9369) |
1 |
The date and time this X-Ray source (see Section C.8.2.2.1.1) was first used in this Multi-energy acquisition. |
>Source End DateTime |
(0018,936A) |
1 |
The date and time this X-Ray source (see Section C.8.2.2.1.1) was last used in this Multi-energy acquisition. |
>Switching Phase Number |
(0018,936B) |
1C |
A number, unique within the sequence, to identify the switching phase. Required if Multi-energy Source Technique (0018,9368) is "SWITCHING_SOURCE". |
>Switching Phase Nominal Duration |
(0018,936C) |
3 |
Duration, in microseconds, that the energy is nominally in the target KV for this switching phase. I.e., the Switching Phase Nominal Duration does not include the Switching Phase Transition Duration (0018,936D). The target KV is the value of KVP (0018,0060) for the item in the CT X-Ray Details Sequence (0018,9325) that identifies the Multi-energy CT Path Index (0018,937A) that corresponds to this X-Ray Source. NoteApplicable if Multi-energy Source Technique (0018,9368) is "SWITCHING_SOURCE". |
>Switching Phase Transition Duration |
(0018,936D) |
3 |
Duration, in microseconds, that the energy has left the target KV for this switching phase, but has not yet reached the target KV for the next phase. The target KV is the value of KVP (0018,0060) for the item in the CT X-Ray Details Sequence (0018,9325) that identifies the Multi-energy CT Path Index (0018,937A) that corresponds to this X-Ray Source. NoteApplicable if Multi-energy Source Technique (0018,9368) is "SWITCHING_SOURCE". |
>Generator Power |
(0018,1170) |
3 |
Power in kW going into the X-Ray generator. |
Each item in the Multi-energy CT X-Ray Source Sequence (0018,9365) might describe either a constant source, a constant source with energy selective filter or one output corresponding to a specific energy of a KV switching source. The Attributes will refer to a source, meaning one item in the Multi-energy CT X-Ray Source Sequence (0018,9365). The Attributes use the phrase "physical X-Ray Source" when it is necessary to refer to the actual device (tube) rather than the current item.
Table C.8.2.2-3 specifies the Attributes of the Multi-energy CT X-Ray Detector Macro, which describe the CT X-Ray Detector(s).
Table C.8.2.2-3. Multi-energy CT X-Ray Detector Macro Attributes
Attribute Name |
Tag |
Type |
Attribute Description |
---|---|---|---|
Multi-energy CT X-Ray Detector Sequence |
(0018,936F) |
1 |
X-Ray Detector (see Section C.8.2.2.1.1) information. One or more Items shall be included in this Sequence. |
>X-Ray Detector Index |
(0018,9370) |
1 |
Identification number of this item in the Multi-energy CT X-Ray Detector Sequence. The number shall be 1 for the first Item and increase by 1 for each subsequent Item. |
>X-Ray Detector ID |
(0018,9371) |
1 |
Identifier of the physical X-Ray detector. This might be the serial number. When a single detector discriminates different energies, the X-Ray Detector ID (0018,9371) will have the same value in different Items of Multi-energy CT X-Ray Detector Sequence (0018,936F). |
>Multi-energy Detector Type |
(0018,9372) |
1 |
Technology used to detect multiple energies. Defined Terms:
|
>X-Ray Detector Label |
(0018,9373) |
3 |
Label of this item in the Multi-energy CT X-Ray Detector Sequence. NoteThe label might be High, Low or some nominal bin energy. |
>Nominal Max Energy |
(0018,9374) |
1C |
Nominal maximum energy in keV of photons that are integrated/counted by the detector (see Section C.8.2.2.1.1). Due to energy resolution limits of the detector, some photons above the nominal maximum may be counted. Required if Multi-energy Detector Type (0018,9372) is PHOTON_COUNTING. May be present otherwise. |
>Nominal Min Energy |
(0018,9375) |
1C |
Nominal minimum energy in keV of photons that are integrated/counted by the detector (see Section C.8.2.2.1.1). Due to energy resolution limits of the detector, some photons below the nominal minimum may be counted. Required if Multi-energy Detector Type (0018,9372) is PHOTON_COUNTING. May be present otherwise. |
>Effective Bin Energy |
(0018,936E) |
3 |
Energy of the heterogeneous (polychromatic) photon beam represented by this detector (see Section C.8.2.2.1.1) calculated as if it were monochromatic. NoteE.g., this can be calculated based on the beam spectrum or derived from the attenuation of phantom measurement. |
Each item in the Multi-energy CT X-Ray Detector Sequence (0018,936F) might describe either an integrating detector or one output corresponding to a specific spectrum of a physical detector like:
One layer of a multi-layer detector.
One energy bin of a photon counting detector.
One energy threshold of a photon counting detector.
The Sequence Attribute descriptions will refer to a detector, meaning one item in the Multi-energy CT X-Ray Detector Sequence (0018,936F). The Attributes use the phrase "physical detector" when it is necessary to refer to the actual device rather than the current item.