Computed Radiography ImageCIOD
CT ImageCIOD
MR ImageCIOD
Nuclear Medicine ImageCIOD
Ultrasound ImageCIOD
Ultrasound Multi-frame ImageCIOD
Secondary Capture ImageCIOD
Multi-frame Single Bit Secondary Capture ImageCIOD
Multi-frame Grayscale Byte Secondary Capture ImageCIOD
Multi-frame Grayscale Word Secondary Capture ImageCIOD
PatientMModule - Patient
Clinical Trial SubjectUModule - Patient
General StudyMModule - Study
Patient StudyUModule - Study
Clinical Trial StudyUModule - Study
General SeriesMModule - Series
Clinical Trial SeriesUModule - Series
General EquipmentUModule - Equipment
SC EquipmentMModule - Equipment
Frame of ReferenceCModule - Frame of Reference
SynchronizationUModule - Frame of Reference
General AcquisitionMModule - Acquisition
General ImageMModule - Image
(0008,0008) Image Type3Code String
(0008,0023) Content Date2CDate
(0008,0033) Content Time2CTime
(0008,2218) Anatomic Region Sequence3Sequence
(0008,2228) Primary Anatomic Structure Sequence3Sequence
(0020,0013) Instance Number2Integer String
(0020,0020) Patient Orientation2CCode String
(0020,0062) Image Laterality3Code String
(0020,4000) Image Comments3Long Text
(0028,0300) Quality Control Image3Code String
(0028,0301) Burned In Annotation3Code String
(0028,0302) Recognizable Visual Features3Code String
(0028,2110) Lossy Image Compression3Code String
(0028,2112) Lossy Image Compression Ratio3Decimal String
(0028,2114) Lossy Image Compression Method3Code String
(0040,9096) Real World Value Mapping Sequence3Sequence
(0088,0200) Icon Image Sequence3Sequence
(0028,0002) Samples per Pixel1Unsigned Short
(0028,0004) Photometric Interpretation1Code String
(0028,0006) Planar Configuration1CUnsigned Short
(0028,0010) Rows1Unsigned Short
(0028,0011) Columns1Unsigned Short
(0028,0034) Pixel Aspect Ratio1CInteger String
(0028,0100) Bits Allocated1Unsigned Short
(0028,0101) Bits Stored1Unsigned Short
(0028,0102) High Bit1Unsigned Short
(0028,0103) Pixel Representation1Unsigned Short
(0028,0106) Smallest Image Pixel Value3US or SS
(0028,0107) Largest Image Pixel Value3US or SS
(0028,1101) Red Palette Color Lookup Table Descriptor1CUS or SS
(0028,1102) Green Palette Color Lookup Table Descriptor1CUS or SS
(0028,1103) Blue Palette Color Lookup Table Descriptor1CUS or SS
(0028,1201) Red Palette Color Lookup Table Data1COther Word String
(0028,1202) Green Palette Color Lookup Table Data1COther Word String
(0028,1203) Blue Palette Color Lookup Table Data1COther Word String
(0028,2000) ICC Profile3Other Byte String
(0028,2002) Color Space3Code String
(7FE0,0010) Pixel Data1OB or OW
(2050,0020) Presentation LUT Shape3Code String
General ReferenceUModule - Image
Image PixelMModule - Image
CineCModule - Image
Multi-frameMModule - Image
Frame PointersUModule - Image
DeviceUModule - Image
Multi-frame Functional GroupsUModule - Image
Multi-frame DimensionUModule - Image
SpecimenUModule - Image
SC ImageUModule - Image
SC Multi-frame ImageMModule - Image
SC Multi-frame VectorCModule - Image
VOI LUTCModule - Image
SOP CommonMModule - Image
Common Instance ReferenceUModule - Image
Frame ExtractionCModule - Image
Multi-frame True Color Secondary Capture ImageCIOD
X-Ray Angiographic ImageCIOD
X-Ray Radiofluoroscopic ImageCIOD
RT ImageCIOD
RT DoseCIOD
RT Structure SetCIOD
RT PlanCIOD
Positron Emission Tomography ImageCIOD
Digital X-Ray ImageCIOD
Digital Mammography X-Ray ImageCIOD
Digital Intra-Oral X-Ray ImageCIOD
RT Beams Treatment RecordCIOD
RT Brachy Treatment RecordCIOD
RT Treatment Summary RecordCIOD
VL Endoscopic ImageCIOD
VL Microscopic ImageCIOD
VL Slide-Coordinates Microscopic ImageCIOD
VL Photographic ImageCIOD
Video Endoscopic ImageCIOD
Video Microscopic ImageCIOD
Video Photographic ImageCIOD
VL Whole Slide Microscopy ImageCIOD
Real-Time Video Endoscopic ImageCIOD
Real-Time Video Photographic ImageCIOD
Dermoscopic Photography ImageCIOD
Grayscale Softcopy Presentation StateCIOD
Color Softcopy Presentation StateCIOD
Pseudo-Color Softcopy Presentation StateCIOD
Blending Softcopy Presentation StateCIOD
Basic Structured DisplayCIOD
XA/XRF Grayscale Softcopy Presentation StateCIOD
Advanced Blending Presentation StateCIOD
Variable Modality LUT Softcopy Presentation StateCIOD
Basic Voice Audio WaveformCIOD
12-Lead ECGCIOD
General ECGCIOD
Ambulatory ECGCIOD
Hemodynamic WaveformCIOD
Basic Cardiac Electrophysiology WaveformCIOD
Arterial Pulse WaveformCIOD
Respiratory WaveformCIOD
General Audio WaveformCIOD
Real-Time Audio WaveformCIOD
Routine Scalp ElectroencephalogramCIOD
ElectromyogramCIOD
ElectrooculogramCIOD
Sleep ElectroencephalogramCIOD
Multi-channel Respiratory WaveformCIOD
Body Position WaveformCIOD
General 32-bit ECGCIOD
Basic Text SRCIOD
Enhanced SRCIOD
Comprehensive SRCIOD
Key Object Selection DocumentCIOD
Mammography CAD SRCIOD
Chest CAD SRCIOD
Procedure LogCIOD
X-Ray Radiation Dose SRCIOD
Spectacle Prescription ReportCIOD
Colon CAD SRCIOD
Macular Grid Thickness and Volume ReportCIOD
Implantation Plan SR DocumentCIOD
Comprehensive 3D SRCIOD
Radiopharmaceutical Radiation Dose SRCIOD
Extensible SRCIOD
Acquisition Context SRCIOD
Simplified Adult Echo SRCIOD
Patient Radiation Dose SRCIOD
Planned Imaging Agent Administration SRCIOD
Performed Imaging Agent Administration SRCIOD
Rendition Selection DocumentCIOD
Enhanced X-Ray Radiation Dose SRCIOD
Enhanced MR ImageCIOD
MR SpectroscopyCIOD
Enhanced MR Color ImageCIOD
Raw DataCIOD
Enhanced CT ImageCIOD
Spatial RegistrationCIOD
Deformable Spatial RegistrationCIOD
Spatial FiducialsCIOD
Ophthalmic Photography 8 Bit ImageCIOD
Ophthalmic Photography 16 Bit ImageCIOD
Stereometric RelationshipCIOD
Hanging ProtocolCIOD
Encapsulated PDFCIOD
Encapsulated CDACIOD
Real World Value MappingCIOD
Enhanced XA ImageCIOD
Enhanced XRF ImageCIOD
RT Ion PlanCIOD
RT Ion Beams Treatment RecordCIOD
SegmentationCIOD
Ophthalmic Tomography ImageCIOD
X-Ray 3D Angiographic ImageCIOD
X-Ray 3D Craniofacial ImageCIOD
Breast Tomosynthesis ImageCIOD
Enhanced PET ImageCIOD
Surface SegmentationCIOD
Color PaletteCIOD
Enhanced US VolumeCIOD
Lensometry MeasurementsCIOD
Autorefraction MeasurementsCIOD
Keratometry MeasurementsCIOD
Subjective Refraction MeasurementsCIOD
Visual Acuity MeasurementsCIOD
Ophthalmic Axial MeasurementsCIOD
Intraocular Lens CalculationsCIOD
Generic Implant TemplateCIOD
Implant Assembly TemplateCIOD
Implant Template GroupCIOD
RT Beams Delivery InstructionCIOD
Ophthalmic Visual Field Static Perimetry MeasurementsCIOD
Intravascular Optical Coherence Tomography ImageCIOD
Ophthalmic Thickness MapCIOD
Surface Scan MeshCIOD
Surface Scan Point CloudCIOD
Legacy Converted Enhanced CT ImageCIOD
Legacy Converted Enhanced MR ImageCIOD
Legacy Converted Enhanced PET ImageCIOD
Corneal Topography MapCIOD
Breast Projection X-Ray ImageCIOD
Parametric MapCIOD
Wide Field Ophthalmic Photography Stereographic Projection ImageCIOD
Wide Field Ophthalmic Photography 3D Coordinates ImageCIOD
Tractography ResultsCIOD
RT Brachy Application Setup Delivery InstructionCIOD
Planar MPR Volumetric Presentation StateCIOD
Volume Rendering Volumetric Presentation StateCIOD
Content Assessment ResultsCIOD
CT Performed Procedure ProtocolCIOD
CT Defined Procedure ProtocolCIOD
Protocol ApprovalCIOD
XA Performed Procedure ProtocolCIOD
XA Defined Procedure ProtocolCIOD
Ophthalmic Optical Coherence Tomography En Face ImageCIOD
Ophthalmic Optical Coherence Tomography B-scan Volume AnalysisCIOD
Encapsulated STLCIOD
Encapsulated OBJCIOD
Encapsulated MTLCIOD
RT Physician IntentCIOD
RT Segment AnnotationCIOD
RT Radiation SetCIOD
C-Arm Photon-Electron RadiationCIOD
Tomotherapeutic RadiationCIOD
Robotic-Arm RadiationCIOD
RT Radiation Record SetCIOD
RT Radiation Salvage RecordCIOD
C-Arm Photon-Electron Radiation RecordCIOD
Tomotherapeutic Radiation RecordCIOD
Robotic-Arm Radiation RecordCIOD
RT Radiation Set Delivery InstructionCIOD
RT Treatment PreparationCIOD
Enhanced RT ImageCIOD
Enhanced Continuous RT ImageCIOD
RT Patient Position Acquisition InstructionCIOD
Microscopy Bulk Simple AnnotationsCIOD
InventoryCIOD
Photoacoustic ImageCIOD
Confocal Microscopy ImageCIOD
Confocal Microscopy Tiled Pyramidal ImageCIOD
Basic DirectoryCIOD

Built with by Innolitics, a team of medical imaging software developers.

Data synced with official DICOM standard on 18 April 2024. The DICOM Standard is under continuous maintenance, and the current official version is available at http://www.dicomstandard.org/current/. DICOM Parts 3, 4, and 6, © NEMA. Please note that the most recent PDF version of the standard is the official reference, and should checked when making technical decisions.

Red Palette Color Lookup Table Descriptor Attribute

Tag(0028,1101)
TypeConditionally Required (1C)
KeywordRedPaletteColorLookupTableDescriptor
Value Multiplicity3
Value RepresentationUS or SS

Specifies the format of the Red Palette Color Lookup Table Data (0028,1201). Required if Photometric Interpretation (0028,0004) has a value of PALETTE COLOR or Pixel Presentation (0008,9205) at the image level equals COLOR or MIXED. See Section C.7.6.3.1.5 for further explanation.

Section C.7.6.3.1.5

C.7.6.3.1.5 Palette Color Lookup Table Descriptor

The three values of Palette Color Lookup Table Descriptor (0028,1101-1104) describe the format of the Lookup Table Data in the corresponding Data Element (0028,1201-1204) or (0028,1221-1223). In this section, the term "input value" is either the Palette Color Lookup Table input value described in the Enhanced Palette Color Lookup Table Sequence (0028,140B) or if that Attribute is absent, the stored pixel value.

The first Palette Color Lookup Table Descriptor value is the number of entries in the lookup table. When the number of table entries is equal to 216 then this value shall be 0. The first value shall be identical for each of the Red, Green, Blue and Alpha Palette Color Lookup Table Descriptors.

The second Palette Color Lookup Table Descriptor value is the first input value mapped. This input value is mapped to the first entry in the Lookup Table Data. All input values less than the first value mapped are also mapped to the first entry in the Lookup Table Data if the Photometric Interpretation is PALETTE COLOR.

Note

In the case of the Supplemental Palette Color LUT, the stored pixel values less than the second descriptor value are grayscale values.

An input value one greater than the first value mapped is mapped to the second entry in the Lookup Table Data. Subsequent input values are mapped to the subsequent entries in the Lookup Table Data up to an input value equal to number of entries + first value mapped - 1, which is mapped to the last entry in the Lookup Table Data. Input values greater than or equal to number of entries + first value mapped are also mapped to the last entry in the Lookup Table Data. The second value shall be identical for each of the Red, Green, Blue and Alpha Palette Color Lookup Table Descriptors.

The third Palette Color Lookup Table Descriptor value specifies the number of bits for each entry in the Lookup Table Data. It shall take the value of 8 or 16. The LUT Data shall be stored in a format equivalent to 8 bits allocated when the number of bits for each entry is 8, and 16 bits allocated when the number of bits for each entry is 16, where in both cases the high bit is equal to bits allocated-1. The third value shall be identical for each of the Red, Green and Blue Palette Color Lookup Table Descriptors.

Note

Some implementations have encoded 8 bit entries with 16 bits allocated, padding the high bits; this can be detected by comparing the number of entries specified in the LUT Descriptor with the actual value length of the LUT Data entry. The value length in bytes should equal the number of entries if bits allocated is 8, and be twice as long if bits allocated is 16.

When the Red, Green, or Blue Palette Color Lookup Table Descriptor (0028,1101-1103) are used as part of the Palette Color Lookup Table Module or the Supplemental Palette Color Lookup Table Module in an Image or Presentation State IOD, the third value shall be equal to 16. When the Alpha Palette Color Lookup Table Descriptor (0028,1104) is used, the third value shall be equal to 8.

When the Red, Green, or Blue Palette Color Lookup Table Descriptor (0028,1101-1103) are used as part of the Palette Color Lookup Table Module in a Color Palette IOD, the 3rd value of Palette Color Lookup Table Descriptor (0028,1101-1103) (i.e, the number of bits for each entry in the Lookup Table Data) shall be 8.

Note

  1. A value of 16 indicates the Lookup Table Data will range from (0,0,0) minimum intensity to (65535,65535,65535) maximum intensity.

  2. Since the Palette Color Lookup Table Descriptor (0028,1101-1104) Attributes are multi-valued, in an Explicit VR Transfer Syntax, only one value representation (US or SS) may be specified, even though the first and third values are always by definition interpreted as unsigned. The explicit VR actually used is dictated by the VR needed to represent the second value, which will be consistent with Pixel Representation (0028,0103).