Defines if the current SOP Instance records a complete treatment delivery of a RT Treatment Fraction.
Enumerated Values:
See Section C.36.20.1.3.
When there is an RT Radiation Record Instance present for all of the RT Radiation Instances for this RT Treatment Fraction and they all have a Treatment Delivery Continuation Flag (300A,0708) of NO and have an RT Treatment Termination Status (300A,0714) of NORMAL the value of RT Radiation RT Treatment Fraction Completion Status (300A,0706) shall be COMPLETE. The expression "all of the RT Radiation Instances" refers to the referenced Instances in the RT Radiation Set, to which this current RT Radiation Record Set refers.
Otherwise the value of RT Treatment Fraction Completion Status (300A,0706) shall be PARTIAL.
Example
RT Radiation Set P contains RT Radiations A and B.
RT Radiation Record Sets created are W, X, Y, and Z.
Table C.36.20-3. Delivery of a Single RT Radiation Set Instance with Partial Treatments
Treatment Session |
RT Radiation Set |
Delivered RT Radiations |
Treatment Delivery Continuation Flag (300A,0708) |
RT Treatment Termination Status (300A,0714) |
RT Radiation Record Set |
RT Treatment Fraction Completion Status (300A,0706) |
Clinical Fraction Number (300A,0705) |
RT Radiation Set Delivery Number (300A,0704) |
---|---|---|---|---|---|---|---|---|
1 |
P |
A |
NO |
NORMAL |
W |
PARTIAL |
1 |
1 |
B |
NO |
ABNORMAL |
||||||
2 |
P |
B |
YES |
NORMAL |
X |
PARTIAL |
1 |
1 |
A |
NO |
NORMAL |
Y |
COMPLETE |
2 |
2 |
||
B |
NO |
NORMAL |
||||||
3 |
P |
A |
NO |
NORMAL |
Z |
COMPLETE |
3 |
3 |
B |
NO |