Tag | (0008,9007) |
---|---|
Type | Required (1) |
Keyword | FrameType |
Value Multiplicity | 4-5 |
Value Representation | Code String (CS) |
Type of Frame. A multi-valued Attribute analogous to Image Type (0008,0008).
Enumerated Values and Defined Terms are the same as those for the four or five values of Image Type (0008,0008), except that the value MIXED is not allowed unless the SOP Class UID is "1.2.840.10008.5.1.4.1.1.2.2" (Legacy Converted). See Section C.8.16.1 and Section C.8.15.2.1.1.
Part of the CT Image Frame Type Functional Group Macro with usage: M
Image Type (0008,0008) and associated Attributes provide a high level description of a multi-frame SOP Instance. Image Type (0008,0008) contains the highest level summary of what is in the SOP Instance.
Frame Type (0008,9007) mirrors the corresponding Image Type Attribute and applies to the frame level.
If more than one value is used by the set of frames for a given Frame Type (0008,9007) Attribute value or associated Attribute value then the corresponding value of Image Type (0008,0008) or associated Attribute shall contain a value of MIXED. This indicates that a mixed set of values exists within the multi-frame SOP Instance.
The value MIXED shall only be used in Image Type (0008,0008) when the corresponding values for the individual frames are not equal. When a value of an Attribute is equal for all frames, the same value shall be used for the corresponding value of Image Type (0008,0008). Values 2 and 3 of Image Type (0008,0008) are an exception to the rule for MIXED: Values 2 and 3 may never have the value of MIXED as described in Section C.8.16.1.2 and Section C.8.16.1.3.
Image Type (0008,0008) and Frame Type (0008,9007) shall consist of four values, except that a fifth value shall be present if Multi-energy CT Acquisition (0018,9361) has a value of YES.
Value 1 of Image Type (0008,0008) and Frame Type (0008,9007) shall use one of the following Enumerated Values from Table C.8-127.
Value 1 of Image Type (0008,0008) and Value 1 of Frame Type (0008,9007) shall not be zero length.
Table C.8-127. Image Type and Frame Type Value 1
Enumerated Value Name |
Enumerated Value Description |
---|---|
ORIGINAL |
An image or frame is original if its pixel data was directly reconstructed from the original data that is obtained from the sensors of the imaging equipment, Image Type (0008,0008) Value 4 is NONE, and Volume Based Calculation Technique (0008,9207) is NONE. Note
|
DERIVED |
An image or frame is derived if its pixel data was calculated from original or other derived pixel data (i.e., it is not original). |
MIXED |
Used only as a value in Image Type (0008,0008) if frames within the SOP Instance contain different values for Value 1 in their Frame Type (0008,9007). |
Value 2 for Image Type (0008,0008) and Frame Type (0008,9007) follows the standard definition and shall have the following Enumerated Value from Table C.8-128.
Value 2 of Image Type (0008,0008) and Value 2 of Frame Type (0008,9007) shall not be zero length.
Table C.8-128. Image Type and Frame Type Value 2
Enumerated Value Name |
Enumerated Value Description |
---|---|
PRIMARY |
Value 3 is an overall representation of the image type. This value may be a summary of several other Attributes or a duplication of one of the other Attributes to indicate the most important aspect of this image. Value 3 Image Flavor is to be used with Value 4 Derived Pixel Contrast to indicate the nature of the image set.
For example Value 3 = DIFFUSION together with Value 4 = NONE indicates that the image set was originally collected for DIFFUSION.
If Value 3 = DIFFUSION together with Value 4 = DIFFUSION this indicates that the object contains DIFFUSION weighted post processed images.
Value 3 of Image Type (0008,0008) shall not be zero length.
Value 3 of Frame Type (0008,9007) may have the same value as found in Value 3 of Image Type (0008,0008), or may have a different value.
The Attribute value may not be MIXED as this value needs to be a summary of the primary purpose of the images, whether the frames have the same value or not.
Table C.8-129 specifies the Defined Terms for Value 3 for Image Type (0008,0008) and Frame Type (0008,9007) that are common to CT and MR. Additional Defined Terms are defined in the modality-specific Module and Macro definitions.
Table C.8-129. Image Type and Frame Type Value 3 Common
Defined Term Name |
Defined Term Description |
---|---|
ANGIO |
Collected for the purpose of angiography. |
CARDIAC |
Images of the heart. |
CARDIAC_GATED |
Cardiac gated images, other than of the heart. |
CARDRESP_GATED |
Cardiac and respiratory gated images. |
DYNAMIC |
An image in which the same anatomical volume is imaged at multiple times in order to capture images of a non-cyclic, time varying event. For example, imaging of the uptake of a tracer or contrast in a specific organ over time. NoteThis is different from gating techniques, in which the same anatomical volume is imaged during some portion of a cyclic event, e.g., inspiration or R-R Interval. |
FLUOROSCOPY |
Real-time collection of single slices (e.g., CT or MR Fluoroscopy). |
LOCALIZER |
Collected for the purpose of planning other images. |
MOTION |
Collected for looking at body motion. |
PERFUSION |
Collected for the purposes of perfusion calculations. |
PRE_CONTRAST |
Collected before contrast was administered. |
POST_CONTRAST |
Collected during or after contrast was administered. |
RESP_GATED |
Respiratory gated images. |
REST |
Cardiac rest image set. |
STATIC |
A group of frames at varying spatial locations acquired at the same time. |
STRESS |
Cardiac stress image set. |
VOLUME |
Set of frames that define a regularly sampled volume. |
NON_PARALLEL |
Set of frames that are not parallel. |
PARALLEL |
Set of frames that are parallel but do not constitute a regularly sampled volume. |
WHOLE_BODY |
A group of frames of the whole body; the frames may be acquired at various times (as distinct from STATIC). |
Value 4 shall be used to indicate derived pixel contrast - generally, contrast created by combining or processing images with the same geometry. Value 4 shall have a value of NONE when Value 1 is ORIGINAL.
Value 4 may have the value QUANTITY if the derived pixel contrast is described in the Quantity Definition Sequence (0040,9220) of the Real World Value Mapping Macro.
If more than one of the following derived types is applicable, then it is up to the generating application to specify the value that best characterizes the derived image.
Value 4 of Image Type (0008,0008) and Value 4 of Frame Type (0008,9007) shall not be zero length unless the SOP Class UID is "1.2.840.10008.5.1.4.1.1.2.2" or "1.2.840.10008.5.1.4.1.1.4.4" or "1.2.840.10008.5.1.4.1.1.128.1" (Legacy Converted).
Table C.8-130 specifies the Defined Terms for Value 4 for Image Type (0008,0008) and Frame Type (0008,9007) that are common to CT and MR. Additional Defined Terms are defined in the modality-specific Module and Macro definitions.
Table C.8-130. Image Type and Frame Type Value 4 Common
Defined Term Name |
Defined Term Description |
---|---|
ADDITION |
Created through Pixel by pixel addition operation. |
DIVISION |
Created through Pixel by pixel division operation. |
MASKED |
Created through Pixel by pixel masking operation. |
MAXIMUM |
Created through Pixel by Pixel maximum operation. |
MEAN |
Created through Pixel by pixel mean operation. |
MINIMUM |
Created through Pixel by Pixel minimum operation. |
MULTIPLICATION |
Created through Pixel by pixel multiplication operation. |
RESAMPLED |
Pixels have been spatially re-sampled, e.g., MPR. |
STD_DEVIATION |
Standard Deviation. |
SUBTRACTION |
Created through Pixel by pixel subtraction operation. |
NONE |
Not a calculated image. |
QUANTITY |
Derived pixel values are a quantity described by Quantity Definition Sequence (0040,9220) of the Real World Value Mapping Macro. |
MIXED |
Used only as value in Image Type (0008,0008) if frames within the image SOP Instance contain different values for value 4 in their Frame Type (0008,9007) Attribute. |
In addition to the requirements specified in Section C.8.16.1 Image Type and Frame Type, the following additional requirements and Defined Terms are specified.
These requirements and Defined Terms are also applicable to Frame Type (0008,9007).
Value 1 of Image Type (0008,0008) and Frame Type (0008,9007) is discussed in Section C.8.16.1.1. No additional requirements or Defined Terms.
Value 2 of Image Type (0008,0008) and Frame Type (0008,9007) is discussed in Section C.8.16.1.2. No additional requirements or Defined Terms.
Table C.8-115 specifies the Defined Terms for CT additional to those defined in Section C.8.16.1.3 for Value 3 for Image Type (0008,0008) and Frame Type (0008,9007).
Table C.8-115. Image Type and Frame Type Value 3 for CT
Defined Term Name |
Defined Term Description |
---|---|
ATTENUATION |
Collected for the purpose of performing attenuation corrections (e.g., PET attenuation correction). |
CARDIAC_CTA |
Images of the heart collected for the purpose of CT Coronary Angiography (CTA). |
CARDIAC_CASCORE |
Images of the heart collected for the purpose of CT Coronary Calcium Scoring. |
REFERENCE |
Collected for anatomical reference for PET or SPECT. |
Table C.8-116 specifies the Defined Terms for CT additional to those defined in Section C.8.16.1.4 for Value 4 for Image Type (0008,0008) and Frame Type (0008,9007).
Table C.8-116. Image Type and Frame Type Value 4 for CT
Defined Term Name |
Defined Term Description |
---|---|
FILTERED |
An image filter has been applied. |
MEDIAN |
Pixel by pixel median. |
ENERGY_PROP_WT |
Image pixels created through proportional weighting of multiple acquisitions at distinct X-Ray energies. |
Table C.8-116b specifies the Defined Terms for CT for Value 5 for Image Type (0008,0008) and Frame Type (0008,9007). Image Type (0008,0008) Value 5 shall be present if Multi-energy CT Acquisition (0018,9361) has a value of YES.
Table C.8-116b. Image Type and Frame Type Value 5 for CT
Defined Term Name |
Defined Term Description |
---|---|
VMI |
Virtual Monoenergetic Image. Each real-world value mapped pixel represents CT Hounsfield units and is analogous to a CT image created by a monoenergetic (of a specific keV value) X-Ray beam. |
MAT_SPECIFIC |
Material-Specific Image. Each real-world value mapped pixel value represents a property of a material such as attenuation, concentration or density. |
MAT_REMOVED |
An image with the attenuation contribution of one or more materials removed. For pixels that did not contain any of the removed material(s), the pixel values are unchanged. |
MAT_FRACTIONAL |
Material-Fractional Image. Each real-world value mapped pixel represents the fraction of a voxel occupied by a material. |
EFF_ATOMIC_NUM |
Effective Atomic Number Image. Each real-world value mapped pixel represents Effective Atomic Number of the materials in the voxel. |
ELECTRON_DENSITY |
Electron Density Image. Each real-world value mapped pixel represents the number of electrons per unit volume or the electron density relative to water. |
MAT_MODIFIED |
Material-Modified Image. CT Image where real-world value mapped pixels have been modified to highlight a certain target material (either by partially suppressing the background or by enhancing the target material), or to partially suppress the target material. |
MAT_VALUE_BASED |
Value-Based Image. CT Image where real-world value mapped pixels represent a certain value for a specified material |
MIXED |
Used only as a value in Image Type (0008,0008) if frames within the image SOP Instance contain different values for Value 5 in their Frame Type (0008,9007). |
The Defined Terms in Table C.8-116b are also used as Defined Terms for Value 4 of the CT Image IOD multi-energy Image Type (see Section C.8.2.1.1.1).