Computed Radiography ImageCIOD
CT ImageCIOD
MR ImageCIOD
Nuclear Medicine ImageCIOD
Ultrasound ImageCIOD
Ultrasound Multi-frame ImageCIOD
Secondary Capture ImageCIOD
Multi-frame Single Bit Secondary Capture ImageCIOD
Multi-frame Grayscale Byte Secondary Capture ImageCIOD
Multi-frame Grayscale Word Secondary Capture ImageCIOD
Multi-frame True Color Secondary Capture ImageCIOD
X-Ray Angiographic ImageCIOD
X-Ray Radiofluoroscopic ImageCIOD
RT ImageCIOD
RT DoseCIOD
RT Structure SetCIOD
RT PlanCIOD
PatientMModule - Patient
Clinical Trial SubjectUModule - Patient
General StudyMModule - Study
Patient StudyUModule - Study
Clinical Trial StudyUModule - Study
RT SeriesMModule - Series
Clinical Trial SeriesUModule - Series
Frame of ReferenceUModule - Frame of Reference
General EquipmentMModule - Equipment
RT General PlanMModule - Plan
RT PrescriptionUModule - Plan
RT Tolerance TablesUModule - Plan
RT Patient SetupUModule - Plan
RT Fraction SchemeUModule - Plan
RT BeamsCModule - Plan
RT Brachy Application SetupsCModule - Plan
(300A,0200) Brachy Treatment Technique1Code String
(300A,0202) Brachy Treatment Type1Code String
(300A,0206) Treatment Machine Sequence1Sequence
(300A,0210) Source Sequence1Sequence
(300A,0230) Application Setup Sequence1Sequence
(300A,0232) Application Setup Type1Code String
(300A,0234) Application Setup Number1Integer String
(300A,0236) Application Setup Name3Long String
(300A,0238) Application Setup Manufacturer3Long String
(300A,0240) Template Number3Integer String
(300A,0242) Template Type3Short String
(300A,0244) Template Name3Long String
(300A,0250) Total Reference Air Kerma1Decimal String
(300A,0260) Brachy Accessory Device Sequence3Sequence
(300A,0280) Channel Sequence1Sequence
(3006,0084) Referenced ROI Number2CInteger String
(300A,00E1) Material ID3Short String
(300A,0110) Number of Control Points1Integer String
(300A,0271) Channel Effective Length3Decimal String
(300A,0272) Channel Inner Length2CDecimal String
(300A,0273) Afterloader Channel ID3Short String
(300A,0274) Source Applicator Tip Length2CDecimal String
(300A,0282) Channel Number1Integer String
(300A,0284) Channel Length2Decimal String
(300A,0286) Channel Total Time1Decimal String
(300A,0288) Source Movement Type1Code String
(300A,028A) Number of Pulses1CInteger String
(300A,028C) Pulse Repetition Interval1CDecimal String
(300A,0290) Source Applicator Number3Integer String
(300A,0291) Source Applicator ID2CShort String
(300A,0292) Source Applicator Type1CCode String
(300A,0294) Source Applicator Name3Long String
(300A,0296) Source Applicator Length1CDecimal String
(300A,0298) Source Applicator Manufacturer3Long String
(300A,029C) Source Applicator Wall Nominal Thickness3Decimal String
(300A,029E) Source Applicator Wall Nominal Transmission3Decimal String
(300A,02A0) Source Applicator Step Size1CDecimal String
(300A,02A1) Applicator Shape Referenced ROI Number3Integer String
(300A,02A2) Transfer Tube Number2Integer String
(300A,02A4) Transfer Tube Length2CDecimal String
(300A,02B0) Channel Shield Sequence3Sequence
(3006,0084) Referenced ROI Number2Integer String
(300A,00E1) Material ID3Short String
(300A,02B2) Channel Shield Number1Integer String
(300A,02B3) Channel Shield ID2Short String
(300A,02B4) Channel Shield Name3Long String
(300A,02B8) Channel Shield Nominal Thickness3Decimal String
(300A,02BA) Channel Shield Nominal Transmission3Decimal String
(300A,02C8) Final Cumulative Time Weight1CDecimal String
(300A,02D0) Brachy Control Point Sequence1Sequence
(300C,000E) Referenced Source Number1Integer String
(300C,0042) Referenced Reference Image Sequence3Sequence
ApprovalUModule - Plan
General ReferenceUModule - Plan
SOP CommonMModule - Plan
Common Instance ReferenceUModule - Plan
Positron Emission Tomography ImageCIOD
Digital X-Ray ImageCIOD
Digital Mammography X-Ray ImageCIOD
Digital Intra-Oral X-Ray ImageCIOD
RT Beams Treatment RecordCIOD
RT Brachy Treatment RecordCIOD
RT Treatment Summary RecordCIOD
VL Endoscopic ImageCIOD
VL Microscopic ImageCIOD
VL Slide-Coordinates Microscopic ImageCIOD
VL Photographic ImageCIOD
Video Endoscopic ImageCIOD
Video Microscopic ImageCIOD
Video Photographic ImageCIOD
VL Whole Slide Microscopy ImageCIOD
Real-Time Video Endoscopic ImageCIOD
Real-Time Video Photographic ImageCIOD
Dermoscopic Photography ImageCIOD
Grayscale Softcopy Presentation StateCIOD
Color Softcopy Presentation StateCIOD
Pseudo-Color Softcopy Presentation StateCIOD
Blending Softcopy Presentation StateCIOD
Basic Structured DisplayCIOD
XA/XRF Grayscale Softcopy Presentation StateCIOD
Advanced Blending Presentation StateCIOD
Variable Modality LUT Softcopy Presentation StateCIOD
Basic Voice Audio WaveformCIOD
12-Lead ECGCIOD
General ECGCIOD
Ambulatory ECGCIOD
Hemodynamic WaveformCIOD
Basic Cardiac Electrophysiology WaveformCIOD
Arterial Pulse WaveformCIOD
Respiratory WaveformCIOD
General Audio WaveformCIOD
Real-Time Audio WaveformCIOD
Routine Scalp ElectroencephalogramCIOD
ElectromyogramCIOD
ElectrooculogramCIOD
Sleep ElectroencephalogramCIOD
Multi-channel Respiratory WaveformCIOD
Body Position WaveformCIOD
General 32-bit ECGCIOD
Basic Text SRCIOD
Enhanced SRCIOD
Comprehensive SRCIOD
Key Object Selection DocumentCIOD
Mammography CAD SRCIOD
Chest CAD SRCIOD
Procedure LogCIOD
X-Ray Radiation Dose SRCIOD
Spectacle Prescription ReportCIOD
Colon CAD SRCIOD
Macular Grid Thickness and Volume ReportCIOD
Implantation Plan SR DocumentCIOD
Comprehensive 3D SRCIOD
Radiopharmaceutical Radiation Dose SRCIOD
Extensible SRCIOD
Acquisition Context SRCIOD
Simplified Adult Echo SRCIOD
Patient Radiation Dose SRCIOD
Planned Imaging Agent Administration SRCIOD
Performed Imaging Agent Administration SRCIOD
Rendition Selection DocumentCIOD
Enhanced X-Ray Radiation Dose SRCIOD
Enhanced MR ImageCIOD
MR SpectroscopyCIOD
Enhanced MR Color ImageCIOD
Raw DataCIOD
Enhanced CT ImageCIOD
Spatial RegistrationCIOD
Deformable Spatial RegistrationCIOD
Spatial FiducialsCIOD
Ophthalmic Photography 8 Bit ImageCIOD
Ophthalmic Photography 16 Bit ImageCIOD
Stereometric RelationshipCIOD
Hanging ProtocolCIOD
Encapsulated PDFCIOD
Encapsulated CDACIOD
Real World Value MappingCIOD
Enhanced XA ImageCIOD
Enhanced XRF ImageCIOD
RT Ion PlanCIOD
RT Ion Beams Treatment RecordCIOD
SegmentationCIOD
Ophthalmic Tomography ImageCIOD
X-Ray 3D Angiographic ImageCIOD
X-Ray 3D Craniofacial ImageCIOD
Breast Tomosynthesis ImageCIOD
Enhanced PET ImageCIOD
Surface SegmentationCIOD
Color PaletteCIOD
Enhanced US VolumeCIOD
Lensometry MeasurementsCIOD
Autorefraction MeasurementsCIOD
Keratometry MeasurementsCIOD
Subjective Refraction MeasurementsCIOD
Visual Acuity MeasurementsCIOD
Ophthalmic Axial MeasurementsCIOD
Intraocular Lens CalculationsCIOD
Generic Implant TemplateCIOD
Implant Assembly TemplateCIOD
Implant Template GroupCIOD
RT Beams Delivery InstructionCIOD
Ophthalmic Visual Field Static Perimetry MeasurementsCIOD
Intravascular Optical Coherence Tomography ImageCIOD
Ophthalmic Thickness MapCIOD
Surface Scan MeshCIOD
Surface Scan Point CloudCIOD
Legacy Converted Enhanced CT ImageCIOD
Legacy Converted Enhanced MR ImageCIOD
Legacy Converted Enhanced PET ImageCIOD
Corneal Topography MapCIOD
Breast Projection X-Ray ImageCIOD
Parametric MapCIOD
Wide Field Ophthalmic Photography Stereographic Projection ImageCIOD
Wide Field Ophthalmic Photography 3D Coordinates ImageCIOD
Tractography ResultsCIOD
RT Brachy Application Setup Delivery InstructionCIOD
Planar MPR Volumetric Presentation StateCIOD
Volume Rendering Volumetric Presentation StateCIOD
Content Assessment ResultsCIOD
CT Performed Procedure ProtocolCIOD
CT Defined Procedure ProtocolCIOD
Protocol ApprovalCIOD
XA Performed Procedure ProtocolCIOD
XA Defined Procedure ProtocolCIOD
Ophthalmic Optical Coherence Tomography En Face ImageCIOD
Ophthalmic Optical Coherence Tomography B-scan Volume AnalysisCIOD
Encapsulated STLCIOD
Encapsulated OBJCIOD
Encapsulated MTLCIOD
RT Physician IntentCIOD
RT Segment AnnotationCIOD
RT Radiation SetCIOD
C-Arm Photon-Electron RadiationCIOD
Tomotherapeutic RadiationCIOD
Robotic-Arm RadiationCIOD
RT Radiation Record SetCIOD
RT Radiation Salvage RecordCIOD
C-Arm Photon-Electron Radiation RecordCIOD
Tomotherapeutic Radiation RecordCIOD
Robotic-Arm Radiation RecordCIOD
RT Radiation Set Delivery InstructionCIOD
RT Treatment PreparationCIOD
Enhanced RT ImageCIOD
Enhanced Continuous RT ImageCIOD
RT Patient Position Acquisition InstructionCIOD
Microscopy Bulk Simple AnnotationsCIOD
InventoryCIOD
Photoacoustic ImageCIOD
Confocal Microscopy ImageCIOD
Confocal Microscopy Tiled Pyramidal ImageCIOD
Basic DirectoryCIOD

Built with by Innolitics, a team of medical imaging software developers.

Data synced with official DICOM standard on 18 April 2024. The DICOM Standard is under continuous maintenance, and the current official version is available at http://www.dicomstandard.org/current/. DICOM Parts 3, 4, and 6, © NEMA. Please note that the most recent PDF version of the standard is the official reference, and should checked when making technical decisions.

Referenced ROI Number Attribute

Tag(3006,0084)
TypeRequired, Empty if Unknown (2)
KeywordReferencedROINumber
Value Multiplicity1
Value RepresentationInteger String (IS)

Uniquely identifies ROI representing the Channel Shield specified by ROI Number (3006,0022) in Structure Set ROI Sequence (3006,0020) in Structure Set Module within RT Structure Set referenced by Referenced RT Structure Set Sequence (300C,0060) in RT General Plan Module. See Section C.8.8.15.2.

Structure Set Module

C.8.8.5 Structure Set Module

A structure set defines a set of areas of significance. Each area can be associated with a Frame of Reference and zero or more images. Information that can be transferred with each region of interest (ROI) includes geometrical and display parameters, and generation technique.

Table C.8-41. Structure Set Module Attributes

Attribute Name

Tag

Type

Attribute Description

Structure Set Label

(3006,0002)

1

User-defined label for Structure Set.

Structure Set Name

(3006,0004)

3

User-defined name for Structure Set.

Structure Set Description

(3006,0006)

3

User-defined description for Structure Set.

Instance Number

(0020,0013)

3

A number that identifies this object Instance.

Structure Set Date

(3006,0008)

2

Date at which the content of the Structure Set was last modified.

Structure Set Time

(3006,0009)

2

Time at which the content of the Structure Set was last modified.

Referenced Frame of Reference Sequence

(3006,0010)

3

Sequence describing Frames of Reference in which the ROIs are defined. One or more Items are permitted in this Sequence. See Section C.8.8.5.1.

>Frame of Reference UID

(0020,0052)

1

Uniquely identifies Frame of Reference within Structure Set.

>RT Referenced Study Sequence

(3006,0012)

3

Sequence of Studies containing Series to be referenced.

One or more Items are permitted in this Sequence.

See Section C.8.8.5.4.

>>Include Table 10-11 “SOP Instance Reference Macro Attributes”

>>RT Referenced Series Sequence

(3006,0014)

1

Sequence describing Series of images within the referenced Study that are used in defining the Structure Set.

One or more Items shall be included in this Sequence.

>>>Series Instance UID

(0020,000E)

1

Unique identifier for the Series containing the images.

>>>Contour Image Sequence

(3006,0016)

1

Sequence of Items describing images in a given Series used in defining the Structure Set (typically CT or MR images).

One or more Items shall be included in this Sequence.

>>>>Include Table 10-3 “Image SOP Instance Reference Macro Attributes”

Source Series Information Sequence

(3006,004C)

3

Information about Image Series that are sources of ROIs in this Instance.

This also includes Series present in the RT Referenced Series Sequence (3006,0014).

The Image Series may have been the source of the definition, but the ROI has been resampled to the Image Series referenced in the Referenced Frame of Reference Sequence (3006,0010).

One or more Items are permitted in this Sequence.

>Modality

(0008,0060)

1

Type of device, process or method that originally acquired or produced the data used to create the Instances in this Series.

>Series Date

(0008,0021)

1

Date the Series started.

>Series Time

(0008,0031)

1

Time the Series started.

>Series Description

(0008,103E)

1

Description of the Series.

>Series Instance UID

(0020,000E)

1

Unique identifier of the Series.

>Series Number

(0020,0011)

1

A number that identifies this Series.

Structure Set ROI Sequence

(3006,0020)

1

ROIs for current Structure Set.

One or more Items shall be included in this Sequence.

>ROI Number

(3006,0022)

1

Identification number of the ROI. The value of ROI Number (3006,0022) shall be unique within the Structure Set in which it is created.

>Conceptual Volume Identification Sequence

(3010,00A0)

3

Identifies the Conceptual Volume which is represented by this ROI.

Only a single Item is permitted in this Sequence.

>>Include Table 10.33-1 “Conceptual Volume Macro Attributes”.

>Referenced Frame of Reference UID

(3006,0024)

1

Uniquely identifies Frame of Reference in which ROI is defined, specified by Frame of Reference UID (0020,0052) in Referenced Frame of Reference Sequence (3006,0010).

>ROI Name

(3006,0026)

2

User-defined name for ROI.

>ROI Description

(3006,0028)

3

User-defined description for ROI.

>ROI Volume

(3006,002C)

3

Volume of ROI (cubic centimeters).

>ROI DateTime

(3006,002D)

3

DateTime when this ROI was created.

>ROI Generation Algorithm

(3006,0036)

2

Type of algorithm used to generate ROI.

Defined Terms:

AUTOMATIC

calculated ROI

SEMIAUTOMATIC

ROI calculated with user assistance

MANUAL

user-entered ROI

>ROI Generation Description

(3006,0038)

3

User-defined description of technique used to generate ROI.

>RT Protocol Code Sequence

(3010,005B)

3

The protocol(s) selected by the RT Physician.

One or more Items are permitted in this Sequence.

See Section C.36.5.1.2.

>>Include Table 8.8-1 “Code Sequence Macro Attributes”

No Baseline CID is defined.

>ROI Creator Sequence

(3006,004D)

3

The person or device that last modified this ROI.

Only a single Item is permitted in this Sequence.

>>Include Table C.17-3b “Identified Person or Device Macro Attributes”

Organizational Role Code Sequence (0044,010A) DCID 9555 “Radiotherapy Treatment Planning Person Role”.

>ROI Derivation Algorithm Identification Sequence

(3006,0037)

3

Software algorithm that derived the ROI.

Only a single Item is permitted in this Sequence.

>>Include Table 10-19 “Algorithm Identification Macro Attributes”.

>Derivation Code Sequence

(0008,9215)

3

A coded description of how this ROI was derived.

One or more Items are permitted in this Sequence.

See Section C.8.8.5.3 for further explanation.

>>Include Table 8.8-1 “Code Sequence Macro Attributes”.

Enumerated Value (113085, DCM, "Spatial resampling").

>Definition Source Sequence

(0008,1156)

3

Instances containing the source of the ROI Contour information.

Only a single Item is permitted in this Sequence.

>>Include Table 10-11 “SOP Instance Reference Macro Attributes”.

>>Referenced Segment Number

(0062,000B)

1C

The value of Segment Number (0062,0004) in the referenced SOP Instance that identifies the segment that is the source of the ROI Contour information.

Required if Referenced SOP Class UID (0008,1150) is Segmentation Storage ("1.2.840.10008.5.1.4.1.1.66.4").

>>Referenced Fiducial UID

(0070,031B)

1C

The value of Fiducials UID (0070,031A) in the referenced SOP Instance that identifies the fiducial that is the source of the ROI Contour information.

Required if Referenced SOP Class UID (0008,1150) is Spatial Fiducials Storage ("1.2.840.10008.5.1.4.1.1.66.2").

Predecessor Structure Set Sequence

(3006,0018)

3

The Structure Set that has been used to derive the current Structure Set.

Only a single Item is permitted in this Sequence.

>Include Table 10-11 “SOP Instance Reference Macro Attributes”


C.8.8.5.1 Frames of Reference

The Referenced Frame of Reference Sequence (3006,0010) describes a set of frames of reference in which some or all of the ROIs are expressed. Since the Referenced Frame of Reference UID (3006,0024) is required for each ROI, each Frame of Reference used to express the coordinates of an ROI shall be listed in the Referenced Frame of Reference Sequence (3006,0010) once and only once.

Note

As an example, a set of ROIs defined using a single image Series would list the image Series in a single Referenced Frame of Reference Sequence (3006,0010) Item, providing the UID for this referenced Frame of Reference (obtained from the source images), and listing all pertinent images in the Contour Image Sequence (3006,0016).

C.8.8.5.2 Frame of Reference Relationship Sequence and Transformation Matrix (Retired)

Retired. See PS3.3-2011.

Note

The concept of definition of registered Frame of References using the Frame of Reference Relationship Sequence (3006,00C0) formerly present in the Standard is retired. The use of Registration IODs is advised since the introduction of Spatial Registration IOD, which is a much stronger and more general concept, and independent from the specifics of RT Structure Sets. Additionally it is of importance that registrations are decoupled from image and segmentation objects.

C.8.8.5.3 ROI Derivation Sequence

If an ROI is created by re-sampling an existing ROI that is either (a) present in a different Frame of Reference (FoR) and has been resampled to align with the image Series referenced by the current structure set, or (b) present in the same FoR, but its contours are referenced to a different image Series, it is useful to be able to identify that the ROI is resampled. The Derivation Code Sequence (0008,9215) shall be used with code (113085, DCM, "Spatial resampling") to indicate the resampled ROI.

C.8.8.5.4 SOP Class UID in RT Referenced Study Sequence

Since RT Referenced Study Sequence (3006,0012) is Type 3, the Attribute may be omitted.

If RT Referenced Study Sequence (3006,0012) is present with an Item, for each Item, the SOP Class UID is that of the Study itself, not of the Instances within the Study. The Detached Study Management SOP Class (Retired) ("1.2.840.10008.3.1.2.3.1") may be used in Referenced SOP Class UID (0008,1150) in lieu of there being any other private or standard Study SOP Class defined.

Note

The SOP Class of an Instance (e.g., of one of the images within the referenced Study) is not used, even if all of the Instances within the Study are of the same Instance SOP Class.

RT General Plan Module

C.8.8.9 RT General Plan Module

Table C.8-45. RT General Plan Module Attributes

Attribute Name

Tag

Type

Attribute Description

RT Plan Label

(300A,0002)

1

User-defined label for treatment plan.

RT Plan Name

(300A,0003)

3

User-defined name for treatment plan.

RT Plan Description

(300A,0004)

3

User-defined description of treatment plan.

Instance Number

(0020,0013)

3

A number that identifies this object Instance.

RT Plan Date

(300A,0006)

2

Date treatment plan was last modified.

RT Plan Time

(300A,0007)

2

Time treatment plan was last modified.

Treatment Protocols

(300A,0009)

3

Planned treatment protocols.

Plan Intent

(300A,000A)

3

Intent of this plan.

Defined Terms:

CURATIVE

Curative therapy on patient

PALLIATIVE

Palliative therapy on patient

PROPHYLACTIC

Preventative therapy on patient

VERIFICATION

Verification of patient plan using phantom

MACHINE_QA

Quality assurance of the delivery machine (independently of a specific patient)

RESEARCH

Research project

SERVICE

Machine repair or maintenance operation

Treatment Site

(3010,0077)

3

A free-text label describing the anatomical treatment site.

Treatment Site Code Sequence

(3010,0078)

3

Coded description of the treatment site.

One or more Items are permitted in this Sequence.

>Include Table 8.8-1 “Code Sequence Macro Attributes”

BCID 4 “Anatomic Region”.

>Treatment Site Modifier Code Sequence

(3010,0089)

3

Coded description of the laterality of the treatment site.

Only a single Item is permitted in this Sequencee.

>>Include Table 8.8-1 “Code Sequence Macro Attributes”

BCID 244 “Laterality”.

RT Plan Geometry

(300A,000C)

1

Describes whether RT Plan is based on patient or treatment device geometry. See Section C.8.8.9.1.

Defined Terms:

PATIENT

RT Structure Set exists

TREATMENT_DEVICE

RT Structure Set does not exist

Referenced Structure Set Sequence

(300C,0060)

1C

The RT Structure Set on which the RT Plan is based.

Only a single Item shall be included in this Sequence.

Required if RT Plan Geometry (300A,000C) is PATIENT.

>Include Table 10-11 “SOP Instance Reference Macro Attributes”

Referenced Dose Sequence

(300C,0080)

3

Related Instances of RT Dose (for grids and named/unnamed point doses).

One or more Items are permitted in this Sequence.

See Note 1.

>Include Table 10-11 “SOP Instance Reference Macro Attributes”

Referenced RT Plan Sequence

(300C,0002)

3

Related Instances of RT Plan.

One or more Items are permitted in this Sequence.

>Include Table 10-11 “SOP Instance Reference Macro Attributes”

>RT Plan Relationship

(300A,0055)

1

Relationship of referenced plan with respect to current plan.

Defined Terms:

PRIOR

plan delivered prior to current treatment

ALTERNATIVE

alternative plan prepared for current treatment

PREDECESSOR

plan used in derivation of current plan

VERIFIED_PLAN

plan that is verified using the current plan. This value shall only be used if Plan Intent (300A,000A) is present and has a value of VERIFICATION

CONCURRENT

plan that forms part of a set of two or more RT Plan Instances representing a single conceptual 'plan', applied in parallel in one treatment phase

Frame of Reference to Displayed Coordinate System Transformation Matrix

(0070,030B)

3

A 4x4 transformation matrix that transforms a coordinate of the Frame of Reference to a displayed coordinate system.

Only rigid transformation matrices are permitted (see definition in Section C.20.2.1.2).

Matrix elements shall be listed in row major order.


Note

  1. An RT Dose IOD referenced within the Referenced Dose Sequence (300C,0080) can be used for storing grid-based (pixel) data, individual dose points (with optional dose point names), isodose curves, and DVHs.

C.8.8.9.1 Referenced Structure Set Sequence

An RT Plan Geometry (300A,000C) of PATIENT shall signify that an RT Structure Set has been defined upon which the plan geometry is based, and this RT Structure Set shall be specified in the Referenced Structure Set Sequence (300C,0060). An RT Plan Geometry (300A,000C) of TREATMENT_DEVICE shall indicate that no patient geometry is available, and that the RT Plan is being defined with respect to the IEC FIXED Coordinate System.

Section C.8.8.15.2

C.8.8.15.2 Referenced ROI Number

The Structure Set ROI shall be used in the RT Brachy Application Setups Module to describe the 3D coordinates of Accessory Devices, Applicators and Channel Shields, but not individual source positions (see Section C.8.8.15.9 and Section C.8.8.15.10).